AT1329 Sample Counter
AT1329 Sample Counter
Sample counter is designed for simultaneous or separate measurement of gross alpha activity and gross beta activity in various samples.
Sample counter can be used to:
- Measure gross alpha and gross beta activity of counting samples based on aerosol analytical filters of АФА type
- Measure gross alpha and gross beta activity in "thick-layer" counting samples prepared from sample material (e.g. by evaporation, or any other method of preparing "thick-layer" sample)
- Measure gross alpha and gross beta activity in "thin-layer" counting samples prepared from sample material (e.g. by electrolytic deposition)
- Measure activity, flux density, external alpha and beta radiation for sources of 1П9, 2П9, 3П9, 1С0, 2С0, 3С0 type
- Control radiation contamination of surfaces by a smear method
- Instrument control and processing of measurement information is realised by external PC
- Custom calibration settings capability
- Selectable units of measurement
- LED stabilization of measurement path
- Passive background radiation protection
- Measurement database capability
- Methodological support of measurements
-
АТ1329 Sample counter (alpha-beta)
-
АТ1329А Sample counter (alpha)
-
АТ1329B Sample counter (beta)
Scintillation detector
|
АТ1329 (α and β channel): Phoswich detector:
Plastic 28 cm2 with ZnS(Ag) layer 28 cm2 АТ1329А (α channel): ZnS(Ag), 28 cm2 АТ1329B (β channel): Plastic, 28 cm2 |
Sensitivity
|
≥0.25 Bq-1·s-1 (to 239Pu alpha radiation)
≥0.30 Bq-1·s-1 (to 90Sr+90Y beta radiation) |
Registered energy range
|
3 – 7 MeV (α channel)
155 keV – 3.5 MeV (β channel) |
Measurement range of gross activity
|
0 01 – 104 Bq (α channel)
0.1 – 104 Bq (β channel) |
Minimum measured activity
(measurement time – 1 h) |
0.02 Bq (α channel of 239Pu)
0.28 Bq (β channel of 90Sr+90Y) |
Background count rate
|
0.001 s-1 (α channel)
0.75 s-1 (β channel) |
Dimensions
|
230х230х290 mm
|
Weight
|
≤22 kg (AТ1329)
≤10 kg (AT1329A) ≤22 kg (AT1329B) |
Design and specifications are subject to change.
For details see our datasheet.
Specifications
Available configurations: |
Scintillation detector: |
Sensitivity: |
Registered energy range |
Measurement range of gross activity: |
Minimum measured activity (1 h measurement time): |
Background count rate: |
Dimensions: |
Weight: |