AT6105 Spectrometric System for Radiation Monitoring
AT6105 Spectrometric System for Radiation Monitoring
- Indication of spectra and dose rate readings by each detection unit on site plan or terrain map
- Identification of source radionuclide composition
- Energy range expandable to 5 MeV
- Hermetically sealed construction (IР68)
- Backup power source
Applications
Number of DUs in the system
|
1 – 32
|
Maximum distance of communication line between DUs and the PC
|
1000 m
|
Identified radionuclides
|
Medical, industrial and natural
(Library of identified radionuclides can be modified) |
PC interface
|
USB / Ethernet / Bluetooth (via interface adapter)
|
Detection units
|
BDKG-201M
|
BDKG-203М
|
BDKG-205M
|
BDKG-211M
|
BDKG-219M
|
Scintillation detector
|
NaI(Tl)
Ø25х16 mm
|
NaI(Tl)
Ø25х40 mm
|
NaI(Tl)
Ø40х40 mm
|
NaI(Tl)
Ø63х63 mm
|
NaI(Tl)
Ø63х160 mm
|
Energy range
|
20 keV – 3 MeV
|
||||
Measurement range of ambient dose equivalent rate
|
50 nSv/h –
1 mSv/h
|
30 nSv/h –
500 µSv/h
|
30 nSv/h –
300 µSv/h
|
30 nSv/h –
120 µSv/h
|
30 nSv/h –
50 µSv/h
|
Limit of intrinsic relative dose rate measurement error
|
±20%
|
||||
Energy dependence relative to 662 keV (137Cs)
|
±15% (in energy range from 50 keV to 3 MeV)
|
||||
Typical energy resolution at 662 keV (137Cs)
|
8.5%
|
8%
|
7.5%
|
7.5%
|
8%
|
Typical sensitivity to 137Cs gamma radiation, cps/(µSv·h-1)
|
165
|
400
|
800
|
2450
|
6000
|
Response time for dose rate change from 0.1 to 1 µSv/h
|
≤2 s
|
||||
Number of ADC channels
|
1024
|
||||
Interface
|
RS485
|
||||
Dimensions
|
Ø63х313 mm
|
Ø63х333 mm
|
Ø63х333 mm
|
Ø90х350 mm
|
Ø90х430 mm
|
Weight
|
1 kg
|
1 kg
|
1 kg
|
2 kg
|
3.3 kg
|
Design and specifications are subject to change.
For details see our datasheet.
Specifications
Detection units: |
Number of DUs in the system: |
Maximum distance of communication line between DUs and the PC: |
PC interface: |
Identified radionuclides: |
Scintillation detector: |
Energy range: |
Measurement range of ambient dose equivalent rate: |
Limit of intrinsic relative dose rate measurement error: |
Energy dependence relative to 662 keV (137Cs): |
Typical energy resolution at 662 keV (137Cs): |
Typical sensitivity to 137Cs gamma radiation: |
Overall dimensions / weight: |