AT5350/1 Standard Dosimeter
AT5350/1 Standard Dosimeter
Delivery set:
- Electrometer measurement unit (Electrometer)
- Optional ionization chambers by PTW-Freiburg (Germany):
- TM23342 Parallel-plane X-ray chamber (0.02 cm3)
- TМ31010 Cylindrical ionization chamber (0.125 cm3)
- TМ30010 Thimble ionization chamber (0.6 cm3)
- TМ23361 Cylindrical ionization chamber (30 cm3)
- TМ32002 Spherical ionization chamber (1000 cm3)
Measurement range:
- Direct current rate
- Electric charge
- Charge by the method of numerical integration of current
|
1·10-15 – 1·10-6 A
1·10-15 – 1·10-8 C
1·10-14 – 1·10-1 C
|
Measurement accuracy
|
≤(0.1 – 0.5)%
|
Measurement range:
- Air kerma rate
- Air kerma
- Air kerma by the method of numerical integration of kerma rate
|
0.4 µGy/min – 10 kGy/min
0.05 µGy – 15 Gy
0.05 µGy – 1.5 MGy
|
Measurement accuracy
|
≤±3%
|
X-ray and gamma radiation energy range
|
8 keV – 1.33 MeV
|
Leakage current
|
1·10-15 A max
|
Integration time
|
99,999 s max
|
Power supply
|
230 VAC, 50 Hz
|
Power consumption
|
12 V·A max
|
Overall dimensions / weight
|
294х112.5х250 mm / 3.8 kg
|
Integrated high voltage ±(1 – 500) V power source for ionization chambers supply with 1 V set-up increments
|
Library of parameters for 20 Ionization chambers
|
Stores up to 500 measurement results
|
Automatic correction of measurement results taking into account the air density for unpressurised chambers based on the entered temperature and pressure values
|
Settable units of measurement (Gy, Sv, R, A, C)
|
RS232C interface and dedicated digital inputs/outputs
|
Design and specifications are subject to change.
For details see our datasheet.
Specifications
Measurement range: |
Measurement accuracy: |
|
Measurement range: |
Measurement accuracy: |
|
X-ray and gamma radiation energy range: |
Leakage current: ≤1·10-15 A |
Integration time: ≤99,999 s |
Power supply: 230 VAC, 50 Hz |
Power consumption: ≤12 V·A |
Overall dimensions / weight: |
|
Integrated high voltage ±(1 – 500) V power source for ionization chambers supply with 1 V set-up increments |
Library of parameters for 20 Ionization chambers |
Stores up to 500 measurement results |
Automatic correction of measurement results taking into account the air density for unpressurised chambers based on the entered temperature and pressure values |
Settable units of measurement (Gy, Sv, R, A, C) |
RS232C interface and dedicated digital inputs/outputs |